Klinisk & analytiskBB-518163598
Veterinary Auto Hematology Analyzer BH-HA310VET
Introduction: The Hematology Analyzer adopts classical resistance method to automatically count a variety of cells, also can classify simply according to the size of white blood cells, to help the lab easily achieve automation. Hematology Analyzer are perfectly in line with modern laboratory requirements in facilitation of operation as well as data management and communication functions.
Specifikationer
- ModelBH-HA310VET
- Throughput60 samples/hour
- PrincipleImpedance for WBC differentiation, WBC/RBC/PLT count, Colorimetric method for HGB
- Channel2
- Parameters21 parameters(including MID#, GRAN#, LYM%, RBC, MCV, MCHC, RDW-CV, MPV, PCT, P-LCC, WBC, LYM#, MID%, GRAN%, HGB, HCT, MCH, RDW-SD, PLT, PDW, P-LCR)
- Histograms3 histograms(including WBC, RBC, PLT)
- Sample VolumeVenous blood: 10μl; Capillary blood: 10μl; Prediluted: 20μl
- CalibrationManual calibration, automatic calibration, and fresh blood calibration
- Counting ModeVenous blood, capillary blood, prediluted blood
- Printing ModelAuto print, manual print
- LIS SystemSupporting Bi-directional LIS
- LanguageChinese, English, Spanish, French
- QCL-J, X-B
- ReagentDiluent 10L, Hemolytic agent 250ml(animal specific), Probe cleaner 50ml
- Date Input9.7-inch touch screen; mouse, keyboard(optional)
- OutputBuilt-in printer, supporting external printers
- Printer Paper57*30mm
- Interface4 USB ports, 1 network port, and 1 RS232 serial port
- StorageAutomatic storage, 100,000 results with histograms(including sample information)
- Information TransmissionBidirectional LIS communication, HL7 and many other transmission protocols
- UnclogReverse high-pressure flush
- Work Temp.10~40°C
- Power SupplyAC100~240V, 50/60Hz
- Relative Humidity≤80%
- Atmospheric Pressure86.0kPa~106.0kPa
- External Size(W*D*H)295*505*430mm
- Package Size(W*D*H)Instrument: 404*564*616mm; Reagent 1: 340*340*340mm; Reagent 2: 200*150*100mm
- Net Weight20kg
- Gross WeightInstrument: 23kg; Reagent 1: 13kg; Reagent 2: 1kg
Detaljer
Application: The Hematology Analyzer can detect abnormalities of healthy animals, assess sick animals and monitor sick animals undergoing treatment by performing routine blood tests. It is suitable for animal experiment center, drug research center, animal and plant quarantine department, zoo, pet hospital, animal husbandry and veterinary system and other institutions. Features: 1. Detection many animal species, with upgrade function. 2. Seven animal detection modes(cow, horse, sheep, goat, rabbit, dog, cat, camel). 3. Automatic detection, eliminating manual error. 4. Abnormal results intuitive prompt, timely and accurate screening out abnormal experimental results. 5. Three detection modes: venous blood, peripheral blood, pre-diluted peripheral blood. 6. 9. 7-inch color touch screen, data status at a glance. 7. One button for maintenance. 8. It can be turned on continuously for 24 hours. 9. Applying cyanide-free hemolytic agent to sample, safe and environmental protection without stimulation. 10. Automatic system diagnosis and maintenance, with a variety of alarm functions. 11. Fully enclosed reagent line, no contamination. 12. Excellent data management and communication function, automatically store 100,000 results. 13. Supporting online remote diagnosis and quality control. The BH-HA310VET hematology analyzer has three detection modes of venous blood, capillary blood, prediluted blood, including 21 detection parameters: MID# GRAN# LYM% RBC MCV MCHC RDW-CV MPV PCT P-LCC WBC LYM# MID% GRAN% HGB HCT MCH RDW-SD PLT PDW P-LCR Technical Parameters: Model BH-HA310VET Throughput 60 samples/hour Principle Impedance for WBC differentiation, WBC/RBC/PLT count, Colorimetric method for HGB Channel 2 Parameters 21 parameters(including MID#, GRAN#, LYM%, RBC, MCV, MCHC, RDW-CV, MPV, PCT, P-LCC, WBC, LYM#, MID%, GRAN%, HGB, HCT, MCH, RDW-SD, PLT, PDW, P-LCR) Histograms 3 histograms(including WBC, RBC, PLT) Sample Volume Venous blood: 10μl; Capillary blood: 10μl; Prediluted: 20μl Calibration Manual calibration, automatic calibration, and fresh blood calibration Counting Mode Venous blood, capillary blood, prediluted blood Printing Model Auto print, manual print LIS System Supporting Bi-directional LIS Language Chinese, English, Spanish, French QC L-J, X-B Reagent Diluent 10L, Hemolytic agent 250ml(animal specific), Probe cleaner 50ml Date Input 9.7-inch touch screen; mouse, keyboard(optional) Output Built-in printer, supporting external printers Printer Paper 57*30mm Interface 4 USB ports, 1 network port, and 1 RS232 serial port Storage Automatic storage, 100,000 results with histograms(including sample information) Information Transmission Bidirectional LIS communication, HL7 and many other transmission protocols Unclog Reverse high-pressure flush Work Temp. 10~40°C Power Supply AC100~240V, 50/60Hz Relative Humidity ≤80% Atmospheric Pressure 86.0kPa~106.0kPa Carry-Over Rate Parameter WBC RBC HGB PLT CV ≤1.5% ≤1.0% ≤1.0% ≤3.0% Background WBC RBC HGB PLT ≤0.5*10 9 /L ≤0.05*10 12 /L ≤2g/L ≤10*10 9 /L Linearity Parameter Measurement Range CV WBC (1~10.00)*10 9 /L (10.1~99.9)*10 9 ±0.5*10 9 /L ±5% RBC (0.3~1.00)*10 12 /L (1.01~7.00)*10 12 /L ±0.5*10 12 /L ±5% HGB (20~70)g/L (71~200.0)g/L ±2.0g/L ±3% PLT (20~100)*10 9 /L (101~999)*10 9 /L ±10.0*10 9 /L ±10% Precision Parameter Repeatability Measurement Range WBC ≤ 4.0% (3.5~9.5)*10 9 /L RBC ≤ 2.0% (3.8~5.8)*10 12 /L HGB ≤ 2.0% (115~175)g/L PLT ≤ 8.0% (125~350)*10 9 /L HCT ≤3.0% 35%~50% External Size(W*D*H) 295*505*430mm Package Size(W*D*H) Instrument: 404*564*616mm; Reagent 1: 340*340*340mm; Reagent 2: 200*150*100mm Net Weight 20kg Gross Weight Instrument: 23kg; Reagent 1: 13kg; Reagent 2: 1kg