Klinisk & analytiskBB-49353905
Auto Chemistry Analyzer BK-400
Introduction: It is used to quantitatively analyze the clinical chemistry of human serum, plasma, urine, cerebrospinal fluid and other samples. Do not use for other purposes.
Specifikationer
- ModelBK-400
- Analysis MethodEnd-point, Fixed-time, Rate(Kinetic), Turbidimetry
- CertificationCE, FDA, ISO9001, ISO13485
- Reagent Positions90 refrigerated reagent positions
- Sample Volume1~70μl
- Reagent Volume10~350μl
- ProbeTeflon coating, with anti-collision function, liquid level detection function
- Probe WashingAutomatic washing interior and exterior
- Reagent Cooling2~8℃
- Cuvettes120 reusable cuvettes, optical length 6mm
- Mixer ProbeIndependent stirring
- WashingAutomatic cuvettes washing
- STAT FunctionYES
- SpectrophotometryPost-spectral spectrophotometry
- Wavelength340~800nm
- Absorbance0~3.5Abs
- Quality ControlReal-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot(2D), Daily QC, Monthly QC
- LIS SystemBi-direction, support HL7 protocol
- InterfaceTCP/IP network interface
- PrinterExternal optional, multiple reporting mode available
- Temperature15~30℃
- Water ConsumptionDeionized water: <10L/h
- Humidity40%~85%
- Net Weight190kg
- Package Size(W*D*H)1358*935*1387mm
- Gross Weight255kg
Detaljer
Features: 1. 400 tests per hour. 2. 90 reagent positions. 3. 120 reaction cuvettes. 4. Built-in sample barcode system optional. 5. 60 sample positions(90 sample positions optional). 6. Probe with anti-collision function, liquid level detection function. Technical Parameters: Model BK-400 Overall Performance Throughput 400 tests/hour Analysis Method End-point, Fixed-time, Rate(Kinetic), Turbidimetry Certification CE, FDA, ISO9001, ISO13485 Sample&Reagent System Sample Positions 60 or 90 sample positions, built-in barcode system available Reagent Positions 90 refrigerated reagent positions Sample Volume 1~70μl Reagent Volume 10~350μl Probe Teflon coating, with anti-collision function, liquid level detection function Probe Washing Automatic washing interior and exterior Reagent Cooling 2~8℃ Reaction System Temperature Control 37±0.1℃ Cuvettes 120 reusable cuvettes, optical length 6mm Mixer Probe Independent stirring Washing Automatic cuvettes washing STAT Function YES Optical System Light Source 12V/20W halogen-tungsten lamp Spectrophotometry Post-spectral spectrophotometry Wavelength 340~800nm Absorbance 0~3.5Abs Calibration&QC Calibration Linear: K factor, 1-point, 2-point and multipoint linear Non-Linear: Spline, Polygon, Logit-4P, Logit-5P Quality Control Real-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot(2D), Daily QC, Monthly QC Data Management Software Windows 7/8/10, 32 or 64 bit LIS System Bi-direction, support HL7 protocol Interface TCP/IP network interface Printer External optional, multiple reporting mode available Working Conditions Power Supply AC220V±10%, 50/60Hz; 110V±10%, 60Hz, 500VA Temperature 15~30℃ Water Consumption Deionized water: <10L/h Humidity 40%~85% Size&Weight External Size(W*D*H) 1170*775*1145mm Net Weight 190kg Package Size(W*D*H) 1358*935*1387mm Gross Weight 255kg